Open time: 8 : 00 AM - 5 : 30 PM (Monday - Saturday)
Available on back-order
Object color can be measured based on spectrophotometric colorimetry determined from the prismatic reflectivity.

Putting value of standard in reflection ratio and chromaticity diagram enables to make Fall/Pass decisions.

|
Measurement wavelength |
380nm – 780nm |
|
Measurement method |
Comparison measurement using reference specimen |
|
Sample N.A.* |
0.12 (using 10x objective lens) |
|
0.24 (using 20x objective lens) |
|
|
* it differs from objective lens’ N.A. |
|
|
Sample W.D. |
10.1 mm (using 10x objective lens) |
|
3.1 mm (using 20x objective lens) |
|
|
Sample curvature radius |
-1R – ∞, +1R – ∞ |
|
Sample measurement range |
Approximately ø60 µm (using 10x objective lens) |
|
Approximately ø30 µm (using 20x objective lens) |
|
|
Measurement repeatability (2σ) |
±0.1 % and less (during 380 nm – 410 nm measurement) |
|
±0.01 % and less (during 410 nm – 700 nm measurement) |
|
|
Display resolution |
1 nm |
|
Measurement time |
A few secs-several secs (it differs from sampling time) |
|
Light source |
Halogen lamp 12 V 100W |
|
Z-direction movement range of the stage |
85mm |
|
PC interface |
USB |
|
Weight |
Unit Approximately 20 kg |
|
Power source for light source: Approximately 3 kg |
|
|
Control box: Approximately 8 kg |
|
|
External dimensions |
Unit: 300 (W) x 550 (D) x 570 (H) mm |
|
Power source for light source: 150 (w) x 250 (D) x 140 (H) mm |
|
|
Control box: 220 (W) x 250 (D) x 140 (H) mm |
|
|
Power Source |
Power source for light source: 100 V (2.8A)/220 V AC |
|
Control box: 100 V (0.2A)/220V AC |
|
|
Environment for use |
Installed at the horizontal plane with no vibration |
|
Temperature: 23 ±5°C |
|
|
Humidity: 60% or less |
|
|
An environment is not subject to air fluctuation |