The AND MX-50 moisture analyzer is a high-performance instrument designed for accurate moisture determination in various samples such as food, pharmaceuticals, chemicals, and industrial materials. Equipped with a 400W halogen lamp heating system and high-resolution weighing technology, the MX-50 ensures fast, stable, and reliable results.
High accuracy: Maximum sample capacity of 51g with 0.001g readability.
Clear moisture display: 0.01% / 0.1% resolution for precise tracking of moisture changes.
Reliable results: Standard deviation as low as ±0.10% (for samples >1g).
20 measurement program memory for repeatable testing and efficiency.
Multiple drying modes: Standard, Quick, Step, and Ramp for versatile applications.
Large VFD display for easy reading of parameters.
RS-232C interface for convenient data transfer to a PC.
GLP/GMP/ISO compliance ensuring international laboratory standards.
Food industry: Moisture analysis of flour, grains, spices, and processed foods.
Pharmaceutical industry: Moisture testing of raw materials and finished products.
Chemicals & materials: Determining moisture in plastics, powders, ceramics, and paper.
Agriculture: Quick testing of moisture in seeds, coffee, pepper, and other crops.
MX-50 | |
Measurement Method | 400W straight halogen lamp heating system with SRA filter and SHS weighing technology |
Max Sample Weight Capacity | 51g |
Weight Resolution | 0.001g |
Moisture Content Display | 0.01% / 0.1% |
Moisture Content Accuracy over 1g (Standard Deviation) | 0.10% |
Moisture Content Accuracy over 5g (Standard Deviation) | 0.02% |
Heating Technology | Halogen lamp (Straight type, 400 Watt max, 5000 hours) |
Drying Temperature (1℃ increment) | 50 – 200℃ |
Memory of Measurement Programs | 20 sets |
Measurement Programs | Standard Mode/Automatic Mode/Quick Mode/Timer Mode/Manual Mode |
Measurement Mode | Moisture content (Wet or Dry base) / Dry content / Ratio / Weight |
Heating Mode | Standard / Quick / Step / Ramp |
Display Type | Large VFD |
Interface | RS-232C standard |
Data Memory Function | 100 |
Operating Temperature | 5-40℃ (41-104℉) less than 85%RH |
GLP/GMP/ISO | Available |
Self Check Function | Standard |
Communication Software | WinCT-Moisture standard |
Sample Pan Size | Ø85mm |
Power | AC 100V to 120V (3A) or AC 200V to 240V (1.5A), 50/60 Hz, Approx. 400W |
Physical Dimension / Weight | 215(W) x 320(D) x 173(H) / Approx. 6kg |
Standard Accessories | Sample Pans (20 for MS/MX/MF 10 for ML), Pan Handles (2 for MS/MX/MF, 1 for ML), Tweezers (for MS/MX/MF), Spoon (for MS/MX/MF), Test Sample (30g of Sodium Tartrate Dihydrate for MS/MX/MF), CD-ROM (WinCT-Moisture for MS/MX, WinCT for MF), Glass Fiber Sheet (for MS/MX/MF), RS-232C Cable (for MS/MX), Display Cover, Dust Cover (for MS/MX/MF), Instruction Manual, Quick Reference Card, Power Cable, Fuse |