The manual line up of the HV-100 Series has the following benefits:

  • 144 mm (5.7 inch) colored touchsreen display (Type A) with higher visibility, easy to operate user interface
  • Digimatic output for report creation.
  • USB stick data export
  • Electronic test force change.
  • Max. specimen height up to 210mm 
  • LED illumination for best image. 
  • High number of accessories
  • Vickers, Knoop, Brinell and KC fracture toughness measurement.
  • Supports Brinelltest up to 62,5KG with an optional test force weight.

Features

Test force:
N: 2,942; 4,903; 9,807; 24,51; 49,03; 98,07; 196,1; 294,2
(kgf): 0,3; 0,5; 1; 2,5; 5; 10; 20; 30
External dimensions
[mm]:
System A: 307(W)×696(D)×786(H)mm System B/C/D: 307(W)×627(D)×875(H)mm
Objective lens:
Up to 3 pcs. mountable (one standard lens 10X mounted as standard)
Force dwell time:
5-999 sec (selectable)
Optical system:
Infinity corrected optical system
Illumination unit:
LED light
Standard:
JIS B 7725, ISO 6507-2
Data output:
RS-232C, Digimatic, USB 2. 0 interface
Indenter approach speed:
60 µm/s, 150 µm/s selectable
Control panel:
Built-in touch panel, 5.7″ Color LCD (HM-110A/120A for System A), Control software (PC for System B/C/D)
Functions:
Calculation of Vickers / Knoop*2 / Brinell*3 hardness, and ceramic fracture toughness based on IF method (JIS R1697), 3 display format (standard, list, simple), GO/NG judgment, test condition guide, curve and user correction, hardness corresponding value, statistics calculation
Main unit mass:
HV-110: Approx. 60kg, HV-120: Approx. 58kg
Output:
Digimatic, serial, USB2.0 series A (for memory)*1, USB2.0 B Type (for system communication)
Resolution of diagonal
length of an indentation:
Objective lens less than 50X: 0.1μm (Objective lens more than 50X: 0.01μm)
Specimen dimensions:
System A: height 210mm, depth 170mm (when using flat anvil)
System B: height 181mm, depth 170mm (when using manual XY stage 50X50mm)
System C: height 172mm, depth 170mm, System D: height 132mm, depth 170mm
Test force control:
Lever method and automatic control (load, duration, unload)
Turret drive:
Motor driven and manual operation
Power supply report:
UN 38.3 Test Summary Report

• Equipped with a high-performance motorized test force switching function and motorized turret mounts. Operates as smoothly as electric models.
• List function makes it easy to perform multi-point and average hardness evaluations, which required effort with conventional manual models. This enables consistency of test results.
• Motorized test force switching and motorized turret mounts are easily controlled via a color touch screen control panel.